Consistency in the Assessment of Dried Blood Spot Specimen Size and Quality in U.K. Newborn Screening Laboratories

Autor: Stuart J. Moat, James R. Bonham, Christine Cavanagh, Margaret Birch, Caroline Griffith, Lynette Shakespeare, Clare Le Masurier, Claire Manfredonia, Beverly Hird, Philippa Goddard, Sarah Smith, Laura Wainwright, Rachel S. Carling, Jennifer Cundick, Fiona Jenkinson, Catherine Collingwood, Nick Flynn, Nazia Taj, Mehdi Mirzazadeh, Tejswurree Ramgoolam, Liz Robinson, Amy Headley, Tessa Morgan, David Elliman, Lesley Tetlow
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: International Journal of Neonatal Screening, Vol 10, Iss 3, p 60 (2024)
Druh dokumentu: article
ISSN: 2409-515X
DOI: 10.3390/ijns10030060
Popis: In 2015, U.K. newborn screening (NBS) laboratory guidelines were introduced to standardize dried blood spot (DBS) specimen quality acceptance and specify a minimum acceptable DBS diameter of ≥7 mm. The UK ‘acceptable’ avoidable repeat rate (AVRR) is ≤2%. To assess inter-laboratory variability in specimen acceptance/rejection, two sets of colored scanned images (n = 40/set) of both good and poor-quality DBS specimens were distributed to all 16 U.K. NBS laboratories for evaluation as part of an external quality assurance (EQA) assessment. The mean (range) number of specimens rejected in the first EQA distribution was 7 (1–16) and in the second EQA distribution was 7 (0–16), demonstrating that adherence to the 2015 guidelines was highly variable. A new minimum standard for DBS size of ≥8 mm (to enable a minimum of six sub-punches from two DBS) was discussed. NBS laboratories undertook a prospective audit and demonstrated that using ≥8 mm as the minimum acceptable DBS diameter would increase the AVRR from 2.1% (range 0.55% to 5.5%) to 7.8% (range 0.55% to 22.7%). A significant inverse association between the number of specimens rejected in the DBS EQA distributions and the predicted AVVR (using ≥8 mm minimum standard) was observed (r = −0.734, p = 0.003). Before implementing more stringent standards, the impact of a standard operating procedure (SOP) designed to enable a standardized approach of visual assessment and using the existing ≥7 mm diameter (to enable a minimum of four sub-punches from two DBS) as the minimum standard was assessed in a retrospective audit. Implementation of the SOP and using the ≥7 mm DBS diameter would increase the AVRR from 2.3% (range 0.63% to 5.3%) to 6.5% (range 4.3% to 20.9%). The results demonstrate that there is inconsistency in applying the acceptance/rejection criteria, and that a low AVVR is not an indication of good-quality specimens being received into laboratories. Further work is underway to introduce and maintain standards without increasing the AVRR to unacceptable levels.
Databáze: Directory of Open Access Journals