Autor: |
Jinane Bakkali Tahiri, Martin Kyndt, Jennifer Dhont, Akos Gulyban, Juliane Szkitsak, Evelien Bogaert, Nick Reynaert, Manuela Burghelea |
Jazyk: |
angličtina |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
Physics and Imaging in Radiation Oncology, Vol 27, Iss , Pp 100475- (2023) |
Druh dokumentu: |
article |
ISSN: |
2405-6316 |
DOI: |
10.1016/j.phro.2023.100475 |
Popis: |
This study aimed to develop and validate a comprehensive, reproducible and automatic 4DCT Quality Assurance (QA) workflow (QAMotion) that evaluates image accuracy across various regular and irregular breathing patterns. Volume and amplitude deviations, CT number accuracy, and spatial integrity were used as evaluation metrics. For repeatability tests, tolerances were respected with a mean CT number deviation |
Databáze: |
Directory of Open Access Journals |