Autor: |
Winter Adele, Morris Meg E, Piriyaprasarth Pagamas, Bialocerkowski Andrea E |
Jazyk: |
angličtina |
Rok vydání: |
2008 |
Předmět: |
|
Zdroj: |
BMC Musculoskeletal Disorders, Vol 9, Iss 1, p 6 (2008) |
Druh dokumentu: |
article |
ISSN: |
1471-2474 |
DOI: |
10.1186/1471-2474-9-6 |
Popis: |
Abstract Background The current investigation examined the inter- and intra-tester reliability of knee joint angle measurements using a flexible Penny and Giles Biometric® electrogoniometer. The clinical utility of electrogoniometry was also addressed. Methods The first study examined the inter- and intra-tester reliability of measurements of knee joint angles in supine, sitting and standing in 35 healthy adults. The second study evaluated inter-tester and intra-tester reliability of knee joint angle measurements in standing and after walking 10 metres in 20 healthy adults, using an enhanced measurement protocol with a more detailed electrogoniometer attachment procedure. Both inter-tester reliability studies involved two testers. Results In the first study, inter-tester reliability (ICC[2,10]) ranged from 0.58–0.71 in supine, 0.68–0.79 in sitting and 0.57–0.80 in standing. The standard error of measurement between testers was less than 3.55° and the limits of agreement ranged from -12.51° to 12.21°. Reliability coefficients for intra-tester reliability (ICC[3,10]) ranged from 0.75–0.76 in supine, 0.86–0.87 in sitting and 0.87–0.88 in standing. The standard error of measurement for repeated measures by the same tester was less than 1.7° and the limits of agreement ranged from -8.13° to 7.90°. The second study showed that using a more detailed electrogoniometer attachment protocol reduced the error of measurement between testers to 0.5°. Conclusion Using a standardised protocol, reliable measures of knee joint angles can be gained in standing, supine and sitting by using a flexible goniometer. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|