Dose audit for patients undergoing two common radiography examinations with digital radiology systems

Autor: Tolga İnal,, Gökçe Ataç
Jazyk: angličtina
Rok vydání: 2014
Předmět:
Zdroj: Diagnostic and Interventional Radiology, Vol 20, Iss 1, Pp 100-104 (2014)
Druh dokumentu: article
ISSN: 1305-3825
1305-3612
DOI: 10.5152/dir.2013.12122
Popis: PURPOSEWe aimed to determine the radiation doses delivered to patients undergoing general examinations using computed or digital radiography systems in Turkey. MATERIALS AND METHODSRadiographs of 20 patients undergoing posteroanterior chest X-ray and of 20 patients undergoing anteroposterior kidney-ureter-bladder radiography were evaluated in five X-ray rooms at four local hospitals in the Ankara region. Currently, almost all radiology departments in Turkey have switched from conventional radiography systems to computed radiography or digital radiography systems. Patient dose was measured for both systems. The results were compared with published diagnostic reference levels (DRLs) from the European Union and International Atomic Energy Agency. RESULTSThe average entrance surface doses (ESDs) for chest examinations exceeded established international DRLs at two of the X-ray rooms in a hospital with computed radiography. All of the other ESD measurements were approximately equal to or below the DRLs for both examinations in all of the remaining hospitals. Improper adjustment of the exposure parameters, uncalibrated automatic exposure control systems, and failure of the technologists to choose exposure parameters properly were problems we noticed during the study. CONCLUSIONThis study is an initial attempt at establishing local DRL values for digital radiography systems, and will provide a benchmark so that the authorities can establish reference dose levels for diagnostic radiology in Turkey.
Databáze: Directory of Open Access Journals