Development and evaluation of a compact gamma camera for radiation monitoring

Autor: Dong-Hee Han, Seung-Jae Lee, Hak-Jae Lee, Jang-Oh Kim, Kyung-Hwan Jung, Da-Eun Kwon, Cheol-Ha Baek
Jazyk: angličtina
Rok vydání: 2023
Předmět:
Zdroj: Nuclear Engineering and Technology, Vol 55, Iss 8, Pp 2873-2878 (2023)
Druh dokumentu: article
ISSN: 1738-5733
DOI: 10.1016/j.net.2023.04.029
Popis: The purpose of this study is to perform radiation monitoring by acquiring gamma images and real-time optical images for 99mTc vial source using charge couple device (CCD) cameras equipped with the proposed compact gamma camera. The compact gamma camera measures 86 × 65 × 78.5 mm3 and weighs 934 g. It is equipped with a metal 3D printed diverging collimator manufactured in a 45° field of view (FOV) to detect the location of the source. The circuit's system uses system-on-chip (SoC) and field-programmable-gate-array (FPGA) to establish a good connection between hardware and software. In detection modules, the photodetector (multi-pixel photon counters) is tiled at 8 × 8 to expand the activation area and improve sensitivity. The gadolinium aluminium gallium garnet (GAGG) measuring 0.5 × 0.5 × 3.5 mm3 was arranged in 38 × 38 arrays. Intrinsic and extrinsic performance tests such as energy spectrum, uniformity, and system sensitivity for other radioisotopes, and sensitivity evaluation at edges within FOV were conducted. The compact gamma camera can be mounted on unmanned equipment such as drones and robots that require miniaturization and light weight, so a wide range of applications in various fields are possible.
Databáze: Directory of Open Access Journals