Autor: |
Jun Kishihara, Shinichi Niwano, MD, Hidehira Fukaya, Ryo Nishinarita, Ai Horiguchi, Hironori Nakamura, Tazuru Igarashi, Naruya Ishizue, Tamami Fujiishi, Tomoharu Yoshizawa, Jun Oikawa, Akira Satoh, Masami Murakami, Junya Ako |
Jazyk: |
angličtina |
Rok vydání: |
2017 |
Předmět: |
|
Zdroj: |
Journal of Arrhythmia, Vol 33, Iss 6, Pp 637-639 (2017) |
Druh dokumentu: |
article |
ISSN: |
1880-4276 |
DOI: |
10.1016/j.joa.2017.05.005 |
Popis: |
Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient. Keywords: Automatic pacing threshold adjustment algorithm, Pacing failure |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|