Pacing failure caused by automatic pacing threshold adjustment system

Autor: Jun Kishihara, Shinichi Niwano, MD, Hidehira Fukaya, Ryo Nishinarita, Ai Horiguchi, Hironori Nakamura, Tazuru Igarashi, Naruya Ishizue, Tamami Fujiishi, Tomoharu Yoshizawa, Jun Oikawa, Akira Satoh, Masami Murakami, Junya Ako
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: Journal of Arrhythmia, Vol 33, Iss 6, Pp 637-639 (2017)
Druh dokumentu: article
ISSN: 1880-4276
DOI: 10.1016/j.joa.2017.05.005
Popis: Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient. Keywords: Automatic pacing threshold adjustment algorithm, Pacing failure