Autor: |
JaeWon Yang, MD, Suhas P. Dasari, MD, Howard A. Chansky, MD, Nicholas M. Hernandez, MD |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Arthroplasty Today, Vol 30, Iss , Pp 101543- (2024) |
Druh dokumentu: |
article |
ISSN: |
2352-3441 |
DOI: |
10.1016/j.artd.2024.101543 |
Popis: |
Dual-mobility bearings are being increasingly utilized in total hip arthroplasty. Contemporary modular designs often feature inner cobalt-chromium liners that are seated in outer titanium acetabular shells. However, mating of these 2 dissimilar metals may lead to complications. We present a case report of a patient who was found to have osteolysis and acetabular protrusio due to an adverse local tissue reaction from corrosion between a cobalt-chromium liner and titanium acetabular shell that appeared to be well-seated clinically and radiographically. This case suggests that implant failure and associated adverse local tissue reactions may occur even in seemingly well-seated modular dual-mobility implants. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|