Autor: |
Hendrik Paysen, James Wells, Olaf Kosch, Uwe Steinhoff, Lutz Trahms, Tobias Schaeffter, Frank Wiekhorst |
Jazyk: |
angličtina |
Rok vydání: |
2018 |
Předmět: |
|
Zdroj: |
AIP Advances, Vol 8, Iss 5, Pp 056712-056712-5 (2018) |
Druh dokumentu: |
article |
ISSN: |
2158-3226 |
DOI: |
10.1063/1.5006391 |
Popis: |
Magnetic Particle Imaging (MPI) is a quantitative imaging modality with promising features for several biomedical applications. Here, we study quantitatively the raw data obtained during MPI measurements. We present a method for the calibration of the MPI scanner output using measurements from a magnetic particle spectrometer (MPS) to yield data in units of magnetic moments. The calibration technique is validated in a simplified MPI mode with a 1D excitation field. Using the calibrated results from MPS and MPI, we determine and compare the detection limits for each system. The detection limits were found to be 5·10-12 Am2 for MPS and 3.6·10-10 Am2 for MPI. Finally, the quantitative information contained in a standard MPI measurement with a 3D excitation is analyzed and compared to the previous results, showing a decrease in signal amplitudes of the odd harmonics related to the case of 1D excitation. We propose physical explanations for all acquired results; and discuss the possible benefits for the improvement of MPI technology. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|