Autor: |
Hajiyev Chingiz |
Jazyk: |
angličtina |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
Metrology and Measurement Systems, Vol 23, Iss 3, Pp 413-424 (2016) |
Druh dokumentu: |
article |
ISSN: |
2300-1941 |
DOI: |
10.1515/mms-2016-0029 |
Popis: |
In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM). As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|