Comparison of intraocular pressure readings with Perkins, Tonopen, iCare 200, and iCare Home to manometry in cadaveric eyes
Autor: | Monica K Ertel, Leonard K Seibold, Jennifer L Patnaik, Malik Y Kahook |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | International Journal of Ophthalmology, Vol 15, Iss 12, Pp 2022-2027 (2022) |
Druh dokumentu: | article |
ISSN: | 2222-3959 2227-4898 |
DOI: | 10.18240/ijo.2022.12.19 |
Popis: | AIM: To compare intraocular pressure (IOP) readings obtained with Perkins tonometry, iCare Home, iCare 200, and Tonopen to IOP readings obtained with the manometer of a perfusion system to assess the accuracy and reproducibility of each method of tonometry at set pressures. METHODS: The IOP of human cadaveric eyes (n=2) was measured using a manometer inserted into the eye through the optic nerve. IOP measurements were obtained using a Perkins tonometer, iCare Home, iCare 200, and Tonopen. These measurements were compared to set point IOP measurements of a manometer to determine accuracy and reproducibility of each device. RESULTS: Mean IOP readings obtained with the Perkins tonometer compared to manometer readings demonstrated a difference of -1.0±5.0 mm Hg (P=0.45), indicating a lower reading on average than manometery although not significant. Mean IOP difference between iCare 200 and manometer was 5.3±2.2 mm Hg (P |