Popis: |
We report on a novel, cost-effective non-dispersive infrared (NDIR) multi-gas sensor aimed at environmental air pollution monitoring. The rugged design of the K96 sensor core combines highest compactness and low-power consumption with our unique multi-channel cell design, featuring the detection of up to three different gases simultaneously, including CO2, CH4, N2O, and H2O. Our sensing platform allows the selection of the target gases as well as the concentration ranges, thus providing highly customizable gas sensor systems targeting application-specific gas monitoring settings. The sensor core comes with an implemented calibration model, and can address in real time any cross-sensitivity between the NDIR gas-sensing channels. We provide an immensely versatile sensing system while ensuring high sensing stability combined with high precision (2 and N2O, 4). The K96 multi-gas sensor core offers a resilient sensor solution for the increasing demand of compact monitoring systems in the field of environmental monitoring at reasonable costs for medium-to-high volumes. |