Skin Dosimetry with EBT3 Radiochromic Film in Radiotherapy of Parotid Cancer

Autor: Mohammad Taghi Bahreyni Toossi, Mahdi Ghorbani, Farideh Khorshidi, Mohammad Mohammadi, Nastaran Mohamadian, Fateme Akbari, Mahdie Dayani
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Journal of Biomedical Physics and Engineering, Vol 11, Iss 5, Pp 573-582 (2021)
Druh dokumentu: article
ISSN: 2251-7200
DOI: 10.31661/jbpe.v0i0.1083
Popis: Background: Skin is a sensitive organ and should be spared in radiotherapy and irradiation of skin in radiotherapy can cause to acute and late skin effects such as erythema, desquamation, epilation, color change, or even necrosis. Objective: The aim of the present study is to do skin dosimetry in radiotherapy of parotid cancer using Gafchromic EBT3 radiochromic film. EBT3 radiochromic films were calibrated in 0.2-5 Gy dose range.Material and Methods: This is an experimental study in the field of radiotherapy physics. Treatment planning was performed on a RANDO phantom for treatment of parotid cancer by a clinical oncologist. Based on the treatment planning, the skin dose at various points in the overlapping region of right anterior-oblique and right posterior-oblique fields were measured using EBT3 radiochromic film. Results: The minimum and maximum skin doses in a fraction (with 2.0 Gy prescribed dose) were 0.50 Gy and 0.97 Gy, respectively. Based on these values, the total skin dose in 30 treatment fractions (for removed tumor) or in 35 treatment fractions (for unremoved tumor) was in the range of 15-33 Gy. Conclusion: Based on the skin dosimetry results of parotid cancer radiotherapy using EBT3 films, it is predicted that there will occur mild skin reactions and these reactions can be neglected due to being mild.
Databáze: Directory of Open Access Journals