Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma

Autor: Kazunori Murata, Laurel Glaser, Mary Nardiello, Lakshmi V. Ramanathan, Dean C. Carlow
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: Data in Brief, Vol 16, Iss , Pp 81-89 (2018)
Druh dokumentu: article
ISSN: 2352-3409
DOI: 10.1016/j.dib.2017.11.006
Popis: The objective of this study was to examine the analytical performance of 14 comprehensive metabolic panel analytes on the Abaxis Piccolo Xpress® Point of Care analyzer in serum, plasma, and whole blood. A method comparison was performed on all three specimen types intended for use on the Piccolo Xpress®: serum, heparinized plasma, and whole blood. This data is also presented in Murata et al. (2015) [1]. This article includes the actual Bland-Altman bias plots of the difference in results obtained for analytes in the comprehensive metabolic panel from the Abaxis Piccolo Xpress and the comparison instrument, the Ortho Vitros. Keywords: Clinical chemistry, Point-of care testing