Autor: |
Andrea Gonzalez-Montoro, Noriel Pavón, Julio Barberá, Neus Cuarella, Antonio J. González, Santiago Jiménez-Serrano, Alejandro Lucero, Laura Moliner, David Sánchez, Koldo Vidal, José M. Benlloch |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
EJNMMI Physics, Vol 11, Iss 1, Pp 1-25 (2024) |
Druh dokumentu: |
article |
ISSN: |
2197-7364 |
DOI: |
10.1186/s40658-024-00674-8 |
Popis: |
Abstract Objective Positron Emission Tomography (PET) is a well-known imaging technology for the diagnosis, treatment, and monitoring of several diseases. Most PET scanners use a Ring-Shaped Detector Configuration (RSDC), which helps obtain homogeneous image quality but are restricted to an invariable Field-of-View (FOV), scarce spatial resolution, and low sensitivity. Alternatively, few PET systems use Open Detector Configurations (ODC) to permit an accessible FOV adaptable to different target sizes, thus optimizing sensitivity. Yet, to compensate the lack of angular coverage in ODC-PET, developing a detector with high-timing performance is mandatory to enable Time-of-Flight (TOF) techniques during reconstruction. The main goal of this work is to provide a proof of concept PET scanner appropriate for constructing the new generation of ODC-PET suitable for biopsy guidance and clinical intervention during acquisition. The designed detector has to be compact and robust, and its requirements in terms of performance are spatial and time resolutions |