Diagnostic value of Pentacam anterior segment analyzer examination parameters on early keratoconus

Autor: Hui-Yu Wang, Shao-Zhen Zhao, Mei-Nan He, Bei Du, Xiao-Jun Dong
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Guoji Yanke Zazhi, Vol 21, Iss 10, Pp 1812-1815 (2021)
Druh dokumentu: article
ISSN: 1672-5123
DOI: 10.3980/j.issn.1672-5123.2021.10.30
Popis: AIM: To analyze the diagnostic value of Pentacam anterior segment analyzer examination parameters on early keratoconus.METHODS: Retrospective control study. Totally 100 patients(124 eyes)with keratoconus who were treated in the hospital between January 2019 and January 2020 were selected as study group. According to the severity of Amsler-Krumeich classification, patients with keratoconus were divided into mild group(51 eyes), moderate group(43 eyes)and severe group(30 eyes), and another 30 normal people with past history of myopia and astigmatism(30 eyes)were selected as control group. The Pentacam anterior segment analyzer parameters [horizontal central curvature of 3mm diameter range(K1), vertical central curvature of 3mm diameter range(K2), maximum refractive power of corneal anterior surface(Kmax), corneal astigmatism(Cyl), corneal thickness at the thinnest(thinnest local), ROC curve was used to analyze the sensitive indicators in the diagnosis of early keratoconus.RESULTS:The K1, K2, Kmax, Cyl, ISV, IVA, KI and ACD in study group were significantly higher than those in control group(P0.85), of which ISV had the highest diagnostic value.CONCLUSION: Pentacam anterior segment analyzer can effectively measure the keratoconus parameters. There are differences in the parameters among patients with different degrees of keratoconus. Kmax, thinnest local, ISV, IVA and KI are sensitive indicators for diagnosing early keratoconus, and Pentacam anterior segment analyzer examination parameters have a higher diagnostic value on early keratoconus.
Databáze: Directory of Open Access Journals