Popis: |
Abstract This study aimed to investigate the effects of topical anesthetic and fluorescein drops on intraocular pressure (IOP), central corneal thickness (CCT) and biomechanical properties as measured by Corvis ST (CST-Oculus; Wezlar, Germany) in healthy eyes. A cross-sectional observational study was conducted on 46 healthy patients. The CST measurements were obtained before and immediately after the instillation of topical anesthetic and fluorescein drops. Pre-post instillation data were statistically analyzed. IOP measurements were compared to Goldmann's Applanation Tonometry (GAT), which was also performed after drops instillation. Biomechanical parameters analyzed included applanation 1 velocity, applanation 2 velocity, applanation 1 time, applanation 2 time, whole eye movement, deflection amplitude, and stiffness parameter at first applanation. A statistically significant difference in IOP, both for non-corrected IOP (IOPnct) and biomechanically corrected IOP (bIOP), was observed before and after the instillation of eyedrops. Despite this statistical significance, the observed difference lacked clinical relevance. The IOPnct demonstrated a significant difference pre and post-anesthetic and fluorescein instillation compared to GAT (14.99 ± 2.27 mmHg pre-instillation and 14.62 ± 2.50 mmHg post-instillation, versus 13.98 ± 2.04 mmHg, with p-values of 0.0014 and 0.0490, respectively). Comparable findings were noted when justaposing bIOP to GAT (14.53 ± 2.10 mmHg pre-instillation and 13.15 ± 2.25 mmHg post-instillation, against 13.98 ± 2.04 mmHg, with p-values of 0.0391 and 0.0022, respectively). Additionally, CCT measurements revealed a statistically significant elevation following the administration of topical anesthetic and fluorescein drops (from 544.64 ± 39.85 µm to 586.74 ± 41.71 µm, p |