Absence of instabilities and intra-prosthetic dislocations at 7 to 11 years following THA using a fourth-generation cementless dual mobility acetabular cup

Autor: Julien Chouteau, Jean-Charles Rollier, Michel P. Bonnin, Mo Saffarini, Luca Nover, Jean-Christophe Chatelet, Laurent Jacquot
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Journal of Experimental Orthopaedics, Vol 7, Iss 1, Pp 1-8 (2020)
Druh dokumentu: article
ISSN: 2197-1153
DOI: 10.1186/s40634-020-00265-3
Popis: Abstract Purpose Dual-mobility (DM) cups are increasingly used in total hip arthroplasty (THA) but there lacks literature on their long-term results. We aimed to investigate outcomes of a fourth-generation cementless DM acetabular cup at 7–11 years. Methods We retrospectively evaluated 240 consecutive hips that received cementless THA using the same dual mobility cup (Novae Sunfit TH) and femoral stem (Corail). Patients were recalled at ≥7 years to collect Oxford hip scores (OHS), Harris hip scores (HHS), and inspect for radiolucent lines and granulomas. Multi-variable analyses were performed to determine whether HHS or OHS were associated with pre- or intra-operative variables. Results At 8.4 ± 0.8 years (range, 7–11), 6 hips were revised (2.5%), 54 deceased (22.5%), and 14 could not be reached (5.8%). Four revisions (2 cup+stem, 2 liners only) were due to sepsis (1.7%), one (cup and stem) for trauma (0.4%), and one (stem) due to aseptic loosening (0.4%). For the remaining 166 hips, HHS was 83.6 ± 13.2 and OHS was 20.3 ± 6.7. Multi-variable analysis confirmed that HHS (β = − 0.38; p = 0.039) and OHS (β = 0.36; p