Popis: | We have performed a study of several cesium oven designs. A comparison between recirculating (or sticking-wall) or collimating (or reemitting-wall) ovens is made in order to extract the most efficient design in terms of beam brightness. Unfortunately, non-reproducible behaviors have been observed, and the most often observed output flux is similar to the sticking-wall case, which is the lowest theoretical value of the two cases, with a beam brightness close to $10^{18}$ at.sr$^{-1}$.s$^{-1}$.cm$^{-2}$. The reason of this universally observed behavior is unclear despite having tested several materials for the collimating tube. Conclusion on possible improved design based on sticking of cesium on several (un)cleaned surface is given. Comment: This article may be downloaded for personal use only. Any other use requires prior permission of the author and AIP Publishing. This article appeared in Rev Sci Instrum 1 April 2022; 93 (4): 043302 and may be found at https://pubs.aip.org/aip/rsi/article/93/4/043302/2848701 |