Spectroscopic performance evaluation and modeling of a low background HPGe detector using GEANT4
Autor: | Thakur, Swati, Devi, Soni, Kaintura, Sanjeet S., Tiwari, Katyayni, Singh, Pushpendra P. |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Druh dokumentu: | Working Paper |
Popis: | Low background gamma spectrometry employing HPGe detectors is a sensitive technique for measuring low-level radioactivity in environmental applications, material screening, and for rare decay searches. This work presents spectroscopic performance evaluation and modelling of a low background measurement setup developed at IIT Ropar in Punjab, India, to measure trace natural radioactive elements, with a particular interest in studying low-level radioactivity in soil and/or rock samples to generate specific inputs for low background experiments. The performance test and characterization of a low background cryocooled HPGe detector with relative efficiency of $\sim$33% have been carried out. An effective detector model has been developed using GEANT4 Monte Carlo simulation to determine the response of the detector over an energy range of 80.9-1408 keV and compared with the experimental performance of the detector. The response of the detector obtained using Monte Carlo simulations agrees reasonably well within 93% level of confidence, indicating only 7% deviation in the comparison. The present setup offers improved detection limits of primordial radionuclides (U/Th and K) to measure radioactive contamination in environmental matrices, which has been used elsewhere [1]. Comment: 17 pages, 24 figures, 6 tables |
Databáze: | arXiv |
Externí odkaz: |