Optical pumping enhancement of a free-induction-decay magnetometer

Autor: Hunter, Dominic, Mrozowski, Marcin S., McWilliam, Allan, Ingleby, Stuart J., Dyer, Terry E., Griffin, Paul F., Riis, Erling
Rok vydání: 2023
Předmět:
Zdroj: Journal of the Optical Society of America B, vol 40, issue 10, pp. 2489-2683 (2023)
Druh dokumentu: Working Paper
DOI: 10.1364/JOSAB.501086
Popis: Spin preparation prior to a free-induction-decay (FID) measurement can be adversely affected by transverse bias fields, particularly in the geophysical field range. A strategy that enhances the spin polarization accumulated before readout is demonstrated, by synchronizing optical pumping with a magnetic field pulse that supersedes any transverse fields by over two order of magnitude. The pulsed magnetic field is generated along the optical pumping axis using a compact electromagnetic coil pair encompassing a micro-electromechanical systems (MEMS) vapor cell. The coils also resistively heat the cesium (Cs) vapor to the optimal atomic density without spurious magnetic field contributions as they are rapidly demagnetized to approximately zero field during spin readout. The demagnetization process is analyzed electronically, and directly with a FID measurement, to confirm that the residual magnetic field is minimal during detection. The sensitivity performance of this technique is compared to existing optical pumping modalities across a wide magnetic field range. A noise floor sensitivity of $238\,\mathrm{fT/\surd{Hz}}$ was achieved in a field of approximately $\mathrm{50\,\mu{T}}$, in close agreement with the Cram\'{e}r-Rao lower bound (CRLB) predicted noise density of $258\,\mathrm{fT/\surd{Hz}}$.
Comment: 10 pages, 7 figures
Databáze: arXiv