Autor: |
Bayer, Fritz M., Jezzard, Peter, Bock, Michael, Smith, Alex K. |
Rok vydání: |
2021 |
Předmět: |
|
Druh dokumentu: |
Working Paper |
DOI: |
10.1002/mrm.28940 |
Popis: |
Purpose: Quantitative magnetization transfer (qMT) imaging can be used to quantify the proportion of protons in a voxel attached to macromolecules. Here, we show that the original qMT balanced steady-state free precession (bSSFP) model is biased due to over-simplistic assumptions made in its derivation. Theory and Methods: We present an improved model for qMT bSSFP, which incorporates finite radio-frequency (RF) pulse effects as well as simultaneous exchange and relaxation. Further, a correction to finite RF pulse effects for sinc-shaped excitations is derived. The new model is compared to the original one in numerical simulations of the Bloch-McConnell equations and in previously acquired in-vivo data. Results: Our numerical simulations show that the original signal equation is significantly biased in typical brain tissue structures (by 7-20 %) whereas the new signal equation outperforms the original one with minimal bias (< 1%). It is further shown that the bias of the original model strongly affects the acquired qMT parameters in human brain structures, with differences in the clinically relevant parameter of pool-size-ratio of up to 31 %. Particularly high biases of the original signal equation are expected in an MS lesion within diseased brain tissue (due to a low T2/T1-ratio), demanding a more accurate model for clinical applications. Conclusion: The improved model for qMT bSSFP is recommended for accurate qMT parameter mapping in healthy and diseased brain tissue structures. |
Databáze: |
arXiv |
Externí odkaz: |
|