Neutron hardness of EJ-276 scintillation material

Autor: Mianowski, S., Brylew, K., Dziedzic, A., Grzenda, K., Karpowicz, P., Korgul, A., Krakowiak, M., Prokopowicz, R., Madejowski, G., Mianowska, Z., Moszynski, M., Szczesniak, T., Ziemba, M.
Rok vydání: 2020
Předmět:
Zdroj: 2020 JINST 15 P10012
Druh dokumentu: Working Paper
DOI: 10.1088/1748-0221/15/10/P10012
Popis: This paper presents the results of the fast neutron irradiation (E$_n$ > 0.5MeV) of an EJ-276 scintillator performed in the MARIA research reactor with fluence up to 5.3$\times$10$^{15}$ particles/cm$^2$. In our work, four samples with size $\phi$25.4~mm$\times$5~mm were tested. The changes in the light yield, emission and absorption spectrum and neutron/gamma discrimination using PuBe source before and after irradiation are presented. The figure of merit in neutron/gamma discrimination based on the charge integration method for different neutron fluences and different short gate integration times are determined.
Comment: 13 pages, 15 figures