Simulation results on a real-time in water tritium monitor

Autor: Azevedo, C. D. R., Baeza, A., Chauveau, E., Corbacho, J. A., Díaz, J., Domange, J., Marquet, C., Martínez-Roig, M., Piquemal, F., Veloso, J. F. C. A., Yahlali, N.
Rok vydání: 2020
Předmět:
Druh dokumentu: Working Paper
DOI: 10.1016/j.nima.2020.164555
Popis: In this work we present simulation results for a modular tritium in-water real-time monitor. The system allows for scalability in order to achieve the required sensitivity. The modules are composed by 340 uncladed scintillating fibers immersed in water and 2 photosensors for light readout. Light yield and Birks' coefficient uncertainties for low energy beta particles is discussed. A study of the detection efficiency according to the fiber length is presented. Discussion on the system requirements and background mitigation for a device with sensitivity of 100\,Bq/L, required to comply with the European directive 2013/51/Euratom, is presented. Due to the low energetic beta emission from tritium a detection efficiency close to 3.3\% was calculated for a single 2\,mm round fiber.
Databáze: arXiv