Improved performance of semiconductor laser tracking frequency gauge
Autor: | Kaplan, D. M., Roberts, T. J., Phillips, J. D., Reasenberg, R. D. |
---|---|
Rok vydání: | 2017 |
Předmět: | |
Zdroj: | JINST 13 (2018) P03008 |
Druh dokumentu: | Working Paper |
DOI: | 10.1088/1748-0221/13/03/P03008 |
Popis: | We describe new results from the semiconductor-laser tracking frequency gauge, an instrument that can perform sub-picometer distance measurements and has applications in gravity research and in space-based astronomical instruments proposed for the study of light from extrasolar planets. Compared with previous results, we have improved incremental distance accuracy by a factor of two, to 0.9 pm in 80 s averaging time, and absolute distance accuracy by a factor of 20, to 0.17 $\mu$m in 1000 s. After an interruption of operation of a tracking frequency gauge used to control a distance, it is now possible, using a nonresonant measurement interferometer, to restore the distance to picometer accuracy by combining absolute and incremental distance measurements. Comment: 8 pages, 4 figures. To appear in JINST. (This version with new figures 3 and 4 with improved graphic quality and performance.) |
Databáze: | arXiv |
Externí odkaz: |