Autor: |
Mykkänen, E., Lehtinen, J. S., Kemppinen, A., Krause, C., Drung, D., Nissilä, J., Manninen, A. J. |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
Review of Scientific Instruments, 87, p. 105111 (2016) |
Druh dokumentu: |
Working Paper |
DOI: |
10.1063/1.4963904 |
Popis: |
We measure the current noise of several cryogenic cables in a pulse tube based dilution refrigerator at frequencies between about 1~mHz and 50~kHz. We show that vibration-induced noise can be efficiently suppressed by using vacuum-insulated cables between room temperature and the 2nd pulse tube stage. A noise peak below 4 fA at the 1.4~Hz operation frequency of the pulse tube, and a white noise density of 0.44 fA/\sqrt{Hz} in the millihertz range are obtained. |
Databáze: |
arXiv |
Externí odkaz: |
|