On-orbit calibration of soft X-ray detector on Chang'E-2 satellite
Autor: | Xiao, Hong, Penga, Wenxi, Wang, Huanyu, Cui, Xingzhu, Guo, Dongya |
---|---|
Rok vydání: | 2015 |
Předmět: | |
Druh dokumentu: | Working Paper |
DOI: | 10.1088/1674-1137/39/10/106002 |
Popis: | X-ray spectrometer is one of the satellite payloads on Chang'E-2 satellite. The soft X-ray detector is one of the device on X-ray spectrometer which is designed to detect the major rock-forming elements within 0.5-10keV range on lunar surface. In this paper, energy linearity and energy resolution calibration is done using a weak Fe55 source, while temperature and time effect is considered not take big error. The total uncertainty is estimated to be within 5% after correction. Comment: 6 pages,6 figure |
Databáze: | arXiv |