Attenuation length measurements of liquid scintillator with LabVIEW and reliability evaluation of the device
Autor: | Gao, L., Yu, BX., Ding, YY., Zhou, L., Wen, LJ., Xie, YG., Wang, ZG., Cai, X., Sun, XL., Fang, J., Xue, Z., Zhang, AW., LÜ, QW., Sun, LJ., Ge, YS., Liu, YB., Niu, SL., Hu, T., Cao, J., LÜ, JG. |
---|---|
Rok vydání: | 2013 |
Předmět: | |
Druh dokumentu: | Working Paper |
DOI: | 10.1088/1674-1137/37/7/076001 |
Popis: | The attenuation length measuring device was constructed by using oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested with a variety of ways, the test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.10 plus or minus 0.35 m where Gd-LS was heavily used in Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment. Comment: 5 pages, 8 figures, 9 conferences |
Databáze: | arXiv |
Externí odkaz: |