[Time needs in evaluating digital thoracic images on the monitor in comparison with alternator]

Autor: Ka, Herrmann, Hm, Bonél, Stäbler A, Matzko M, Waggershauser T, Sittek H, Mf, Reiser
Jazyk: němčina
Rok vydání: 2001
Předmět:
Zdroj: Europe PubMed Central
ISSN: 0035-7820
Popis: Evaluation of time efficiency in softcopy reading versus hardcopy reading of digital chest x-rays.130 normal and pathologic chest x-rays in two plains were analyzed by 4 experienced radiologist at both a digital workstation and the light box. Reading time and switch time between two patient folders were measured as well as the frequency of post-processing at the monitor.Reading time at the workstation slightly exceeded reading time at the light box; differences were not statistically significant. Post-processing (frequency between 2% and 83%) did not significantly prolong reading time. Mean switch time between two patient folders was 4.3 sec at the workstation and 13.7 sec at the light box.As compared to hardcopy reading, softcopy reading of digital chest x-rays does not significantly increase reading time. Switch time between patient folders can be reduced at the workstation by a factor of two to four.