Assessment of Head Wear More Than Ten Years after Total Hip Arthroplasty: 22-mm Zirconia VS. Metal Heads

Autor: Atsushi, Inoue, Koji, Asaumi, Hirosuke, Endo, Kazuo, Fujiwara, Shigeru, Mitani, Toshifumi, Ozaki
Jazyk: angličtina
Rok vydání: 2006
Předmět:
Zdroj: Acta Medica Okayama. 60(6):311-318
ISSN: 0386-300X
Popis: The present retrospective study assessed radiographs to determine socket wear in total hip arthroplasty (THA) with 22-mm zirconia or COP (Cobalt-Chrome alloy rich in Cobalt and Phosphorous) heads, and in cemented stems at more than 10 years after operation. Sockets of ultra high molecular weight polyethylene were used in each of two THA groups (13 hips each) in a clinical trial in our hospital between 1989 and 1990. Three observers carried out masked assessments of the radiographs. Upon fi nal examination, there was no remarkable loosening in the zirconia or COP group, and no case had required revision surgery as of 2005. There was a statistically signifi cant diff erence between the 2 groups in average annual linear wear, at 0.093 mm/year and 0.046 mm/year in the zirconia and COP groups, respectively. Volume wear and average annual volume wear were also signifi cantly greater in the zirconia group despite its superior mechanical strength and toughness in vitro. Our present fi ndings do not confi rm early expectations of lower wear in long-term results of 22-mm zirconia femoral heads used in THA.