A multicenter validation of an active contour-based left ventricular analysis technique

Autor: M J, Graves, E, Berry, A A, Eng, M, Westhead, R T, Black, D J, Beacock, S, Kelly, P, Niemi
Rok vydání: 2000
Předmět:
Zdroj: Journal of magnetic resonance imaging : JMRI. 12(2)
ISSN: 1053-1807
Popis: Quantitative analysis of functional cardiac magnetic resonance (MR) images has been limited by the lack of well-validated, semiautomatic, methods for rapid analysis. We describe the evaluation of a DICOM-compatible PC-based parallel-processing tool, for cardiac magnetic resonance analysis (CAMRA), which supports semiautomatic image mensuration using an active contour model-based algorithm. The CAMRA software was used to analyze data from 12 patients in a multicenter acquisition and analysis trial to compare semiautomatic contour detection with manual planimetry of the left ventricular endocardium from short-axis, breath-held, cine gradient-echo images. There was excellent agreement between the manual and semiautomatic measurements of global left ventricular function, with no significant (P = 0.32) difference in the determination of ejection fraction (-0.9 +/- 3.1% [mean difference +/- 1 standard deviation]). There was no significant interobserver difference in the semiautomatically measured ejection fraction. Additionally, a single observer completed the analysis on data from 30 patients and found no significant (P = 0.05) difference in the determination of ejection fraction (-1.3 +/- 3.5% [mean difference +/- 1 standard deviation]). The CAMRA software demonstrates the capability for the reproducible evaluation of global left ventricular function in cardiac patients, with adequate interobserver reproducibility for use in multicenter trials.
Databáze: OpenAIRE