Autor: |
P, Knesewitsch, R, Walser, R, Kantlehner, W, Münzing, K, Hahn |
Jazyk: |
němčina |
Rok vydání: |
1996 |
Předmět: |
|
Zdroj: |
Nuklearmedizin. Nuclear medicine. 35(3) |
ISSN: |
0029-5566 |
Popis: |
In this study our first clinical experiences with simultaneous transmission and emission acquisition in 201 TI myocardial SPECT (T/E-SPECT) are discussed.The non-uniform attenuation (AK) was carried out with a triple-head camera (PRISM 3000, Picker Inc.) correction equipped with fanbeam collimators. A line source of 750 MBq 99mTc was used to construct the transmission profile. Prior to investigation patients got 80-120 MBq 201TI-chloride intravenously injected.The study comprises the evaluation of 40 patients, derived from the clinical routine. The investigation followed an usual one day protocol. Our results using T/E-SPECT reveal an almost equilibrated activity distribution between anterior and posterior myocardial wall.For this reason it is to be expected that T/E-SPECT provides more reliable information about the posterior myocardial wall, than the usual SPECT technique without attenuation correction. |