This study was undertaken to evaluate a quick, reliable and cheap method of digitizing analog scintigrams. 40 whole-body bone scintigrams were obtained simultaneously in analog and genuine digital format. The analog scans on x-ray film were then digitized secondarily by three different methods: 300 dpi flat-bed scanning, high-resolution camera scanning and camcorder recording. A simple exposure approach using a light box, a cheap camcorder, a PC and image grabber hard- and software proved to be optimal. Visual interpretation showed no differences in clinical findings when comparing the analog images with their secondarily digitized counterparts. To test the possibility of quantification, 126 equivalent ROIs were drawn both in the genuine digital and the secondarily digitized images. Comparing the ROI count to whole-body count percentage of the corresponding ROIs showed the correlation to be linear. The evaluation of phantom studies showed the linear correlation to be true within a wide activity range. Thus, secondary digitalization of analog scintigrams is an easy, cheap and reliable method of archiving images and allows secondary digital quantification. |