Autor: |
W E, Frank, B E, Llewellyn, P A, Fish, A K, Riech, T L, Marcacci, D W, Gandor, D, Parker, R R, Carter, S M, Thibault |
Rok vydání: |
2001 |
Předmět: |
|
Zdroj: |
Journal of forensic sciences. 46(3) |
ISSN: |
0022-1198 |
Popis: |
According to TWGDAM guideline 4.5 (1), prior to implementing a new DNA analysis procedure or an existing DNA analysis procedure developed by another laboratory, the forensic laboratory must first demonstrate reliability of the procedure in-house. Seven phases were designed to validate the use of the AmpFlSTR Profiler Plus PCR Amplification Kit, as well as the PE Applied Biosystems 310 Genetic Analyzer. This report summarizes the results obtained for each of the seven phases of the validation study which included the following evaluations: polymer, reproducibility, sensitivity, stutter, precision, mixtures and nonprobative casework. |
Databáze: |
OpenAIRE |