[Simultaneous determination of vitamin B1 and vitamin C in aqueous solution with piezoelectric crystal quartz sensors]

Autor: W Z, Wei, C W, Hu, K M, Wei, L H, Nie, S Z, Yao
Rok vydání: 1993
Předmět:
Zdroj: Yao xue xue bao = Acta pharmaceutica Sinica. 28(10)
ISSN: 0513-4870
Popis: According to the response properties of piezoelectric crystal quartz sensors to the solution conductivity the frequency shift response was derived to show a linear dependence on the concentrations of vitamin B1 (VB1) and vitamin C (VC). This was experimentally verified and a calibration model for simultaneous determinations of the two vitamins was established. The usefulness of the technique was evaluated by quantitation of mixtures of unknown composition using common multiple linear regression (MLR) and partial least squares (PLS). The average relative standard deviations for six samples were 4.46% for VB1 and 6.63% for VC with MLR; and 1.97% for VB1 and 2.74% for VC with PLS, respectively.
Databáze: OpenAIRE