Autor: |
Hitoshi, Nishihara, Masahiko, Ayaki, Tomiko, Watanabe, Takeo, Ohnishi, Toshiyuki, Kageyama, Shigeo, Yaguchi |
Rok vydání: |
2004 |
Předmět: |
|
Zdroj: |
Nippon Ganka Gakkai zasshi. 108(3) |
ISSN: |
0029-0203 |
Popis: |
To compare the long-term clinical and experimental results of soft acrylic intraocular lenses(IOLs) manufactured by the lathe-cut(LC) method and by the cast-molding(CM) method.This was a retrospective study of 20 patients(22 eyes) who were examined in a 5- and 7-year follow-up study. Sixteen eyes were implanted with polyacrylic IOLs manufactured by the LC method and 6 eyes were implanted with polyacrylic IOLs manufactured by the CM method. Postoperative measurements included best corrected visual acuity, contrast sensitivity, biomicroscopic examination, and Scheimpflug slit-lamp images to evaluate surface light scattering. Scanning electron microscopy and three-dimensional surface analysis were conducted.At 7 years, the mean visual acuity was 1.08 +/- 0.24 (mean +/- standard deviation) in the LC group and 1.22 +/- 0.27 in the CM group. Surface light-seatter was 12.0 +/- 4.0 computer compatible tapes(CCT) in the LC group and 37.4 +/- 5.4 CCT in the CM group. Mean surface roughness was 0.70 +/- 0.07 nm in the LC group and 6.16 +/- 0.97 nm in the CM group.Acrylic IOLs manufactured by the LC method are more stable in long-termuse. |