Autor: |
F S, Apple, A J, Maturen, R E, Mullins, P C, Painter, M S, Pessin-Minsley, R A, Webster, J, Spray Flores, R, DeCresce, D J, Fink, P M, Buckley, J, Marsh, V, Ricchiuti, R H, Christenson |
Rok vydání: |
1999 |
Předmět: |
|
Zdroj: |
Clinical chemistry. 45(2) |
ISSN: |
0009-9147 |
Popis: |
We evaluated the AxSYM troponin I (cTnI) immunoassay for assisting in the detection of acute myocardial infarction (AMI). At four sites, the total imprecision (CV) over 20 days was 6.3-10.2%. The minimum detectable concentration was 0.14 +/- 0.05 microgram/L. Comparison of cTnI measurements between the AxSYM and Stratus (n = 406) over the dynamic range of the AxSYM assay demonstrated good correlation, r = 0.881, with a proportional bias: AxSYM cTnI = 3.50(Stratus cTnI) - 1. 10. The confidence intervals (95%) for the slope and intercept were 3.39-3.64 and -1.32 to -0.95, respectively. The expected cTnI concentration in healthy individuals was/=0.5 microgram/L, whereas the ROC curve-determined cutoff for AMI was 2.0 microgram/L. This gave a diagnostic sensitivity of 91.8% and specificity of 92.4% when tested in serial samples collected within 24 h of admission in 633 patients presenting with chest pain, of which 122 had an AMI. The concordances of the AxSYM cTnI with the Stratus cTnI, OPUS cTnI, and Access cTnI were 95.3%, 95.1%, and 94.3%, respectively, from patients with suspected AMI. The AxSYM cTnI demonstrated excellent clinical specificity,/=96%, in skeletal muscle injury, chronic renal disease, and same-day noncardiac surgery patients. |