Autor: |
Charlotte, Struwe, Sebastian G, Walter, Claudia, Druschel, Rahel, Bornemann, Milena, Ploeger, Sebastian, Koob, Richard, Placzek |
Rok vydání: |
2020 |
Předmět: |
|
Zdroj: |
Journal of Materials Science. Materials in Medicine |
ISSN: |
1573-4838 |
Popis: |
The aim of this study is to compare biomechanical features of different devices used in clinical routine for temporary epiphysiodesis (eight-Plate® and FlexTackTM). The tested implants were divided into four different groups (eight-Plate® vs. FlexTackTM for lateral and anterior implantation) á 10 samples for testing implanted eight-Plate® vs. FlexTackTM in fresh frozen pig femora for maximum load forces (Fmax) and axial physis distance until implant failure (lmax). A servo hydraulic testing machine (858 Mini Bionix 2) was used to exert and measure reproducible forces. Statistical analyses tested for normal distribution and significant (p |
Databáze: |
OpenAIRE |