Popis: |
Screw-fixated and hydroxyapatite-coated press-fit cups were studied using radiostereometry in 29 revision and 14 primary arthroplasties. The acetabular defects in the revision cases varied from none to type 3 (wall defects) according to Gustilo-Pasternak. Morsellized allograft was used in 25 revisions. Nine of these cups rested on less than 50% living bone. After 2 years, the mean migration in the revised group reached 0.36, 0.21, and 0.49 mm in the horizontal, longitudinal, and anteroposterior (AP) directions. The mean rotations varied between 0.5 degrees and 0.7 degrees depending on direction. The primary implants displayed smaller mediolateral migration and AP tilt. The mean proximal wear rate for the whole group was 0.11 mm/y. A central gap on the postoperative AP view implied less migration. The size of the preoperative bone defects or amount of bone-graft used had no influence on the migration. Despite extensive use of morsellized allograft, this implant displayed the smallest migration so far reported in revision hip arthroplasty. |