Autor: |
Shih-Ming, Hsu, Chien-Yi, Yeh, Tien-Chi, Yeh, Ji-Hong, Hong, Annie Y H, Tipton, Wei-Li, Chen, Shung-Shung, Sun, David Y C, Huang |
Rok vydání: |
2009 |
Předmět: |
|
Zdroj: |
Medical physics. 35(12) |
ISSN: |
0094-2405 |
Popis: |
High dose rate brachytherapy (HDR-BT) is one of the many modalities for prostate cancer treatment. Due to the nature of HDR-BT, in vivo dosimetry is feasible and can be used to verify consistent dose delivery. In order to validate a dose verification system for HDR-BT prostate cancer treatment, a radiophotoluminescent glass dosimeter (RPLGD) was used and the measurements were compared with those from a thermoluminescent dosimeter. The RPLGD shows many advantages in HDR-BT dose measurement, such as repeatability, stability, and small effective size. These advantages make the RPLGD a superior option for use as a dosimeter in HDR-BT. The results described here show that the difference between the measured dose and the treatment planned dose is less than 5%. A Monte Carlo simulation for the dose was performed using Monte Carlo N -particle to investigate position error. This study concludes that the RPLGD is a promising and reliable dosimeter for HDR-BT in vivo dosimetry with clinically acceptable accuracy. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|