[Comparative test on puncture coring of two different needles used with the implantable drug-supplying device]

Autor: Min, Wan, Ping, Wu, Jinzi, Song, Xin, Yu, Pengtao, Mou
Rok vydání: 2011
Předmět:
Zdroj: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation. 34(6)
ISSN: 1671-7104
Popis: A comparison test on the number of puncture coring is conducted using the normal needle and the Huber needle in the injection area of the implantable drug-supplying device separately. The result indicates that the number of coring using the Huber needle is much less than that using the normal needle during the puncturing. So it is suggested to popularize the Huber needle in the drug transfusion of the implantable drug-supplying device.