Totally subcutaneous implantable cardioverter defibrillator with an alternative, right parasternal, electrode placement

Autor: Sven, Zumhagen, Andrew A, Grace, Stephen, O'Connor, Andreas, Löher, Julia, Köbe, Lars, Eckardt, Eric, Schulze-Bahr
Rok vydání: 2011
Předmět:
Zdroj: Pacing and clinical electrophysiology : PACE. 35(9)
ISSN: 1540-8159
Popis: The totally subcutaneous implantable cardioverter-defibrillator (S-ICD) is an entirely novel defibrillation device that avoids the direct contact of device electrodes with the heart and the cardiovascular system. Here, we present a particular case of a young woman with congenital long-QT syndrome in which we implanted the electrode alternatively, right parasternally. This decision was based on the thoracic anatomy of the patient and on findings of a model of S-ICD electrodes in an adult torso. In conclusion, in some patients an alternative subcutaneous electrode position may be carefully considered but should not be taken to outweigh the standard left-sided placement.