Popis: |
Variation in body position has been shown to affect respiratory function in adults and neonates with and without respiratory illness. At present it remains unclear why respiratory function should be affected by different body positions. We hypothesized that the effect of body weight on the relatively compliant chest wall of the newborn infant in the prone position would cause a reduction in functional residual capacity (FRC) and a compensatory improvement in ventilation/perfusion matching as measured by effective pulmonary blood flow. To evaluate this, a paired crossover study was performed on 12 normal newborn infants. The inert gas (argon) rebreathing method adapted for neonates was used to measure FRC. Simultaneously effective pulmonary blood flow (Qpeff) was determined using Freon 22 and a mass spectrometer with computerized analysis. The babies were studied in three different positions in random order: prone, supine and right lateral decubitus. The means (95% confidence intervals) of the three groups of FRC were 23.8 (19.2 to 28.4), 23.8 (20.2 to 27.5), and 24.3 (19.5 to 29.2) ml/kg, respectively (P = 0.59) and for Qpeff were 104 (91 to 116), 108 (95 to 122), 109 (97 to 122) ml/ kg-min, respectively (P = 0.11). Thus no significant differences were demonstrated. In nine of the babies, a repeat supine measurement was taken at the end of the study to assess repeatability of the method. In these nine babies alone the results were 22.7 (19.1 to 26.3) and 22.1 (18.6 to 25.6) ml/kg for FRC, and 102 (89 to 116) and 98 (90 to 107) ml/kg-min for Qpeff. The coefficients of repeatability were 4.7 ml/kg for FRC (21%) and 30 ml/kg-min for Qpeff (30%). |