Echocardiographic features of aortic ball valve prosthesis malfunction

Autor: M R, Weisburst, A K, Singh, R S, Riley
Rok vydání: 1982
Předmět:
Zdroj: Clinical cardiology. 5(11)
ISSN: 0160-9289
Popis: A case of aortic ball valve prosthesis malfunction is described in which the poppet became alternately stuck in the open and closed position. The patient experienced chest pain followed by pulmonary edema and cardiac arrest. Malfunction of prosthetic valve was diagnosed on echocardiogram and cardiopulmonary resuscitation was carried out until a Bjork-Shiley valve could be inserted in place of the faulty prosthesis. At the time of the operation, poppet migration had occurred and the poppet could not be found. Subsequent Bjork-Shiley aortic valve prosthesis dysfunction was suggested by variation in the intensity of the aortic opening sound and in the duration of the systolic ejection period. Fluoroscopy revealed the missing poppet in the left ventricle. Following surgical removal of the poppet, "normal" Bjork-Shiley valve function was restored.