Saliva samples can be used as an alternative fluid for against hepatitis C virus (anti-HCV) detection owing to the ease of collection and excellent acceptability. This study was conducted to optimize a commercial enzyme immunoassay (EIA) to detect anti-HCV in saliva samples.Ninety-six individuals donated paired serum and saliva samples that were obtained, using a commercial device (Salivette) and spitting into a sterile container. Initially, elution buffer for the Salivette samples, sample volume, incubation time and temperature, and two different anti-HCV EIAs were evaluated. Using the optimized assay, three methods for cut-off calculation were also evaluated.A 20-fold increase in the sample volume for both collection methods was needed. Moreover, the Radim assay was the most appropriate assay for anti-HCV detection in saliva samples, and the quality parameters were increased when a ROC curve was used to determine the cut-off value. Using this optimized assay, the sensitivities, specificities, accuracies, positive and negative predictive values were above 90% for saliva obtained using both the Salivette and spitting methods. Using this assay, discordant false-negative results were obtained for only two Salivette samples and five spitting samples. The concordance kappa was 93% for the Salivette method and 86.1% for the spitting method, demonstrating excellent performance.Saliva samples obtained for both methods can be employed for anti-HCV detection among HCV-infected or HCV-suspected cases, but several modifications must be performed on commercial EIAs to obtain good results. Moreover, samples obtained with commercial devices are more appropriate for anti-HCV detection in saliva samples. |