[The 'high concentration' oxygen mask: a dangerous defect]

Autor: J J, Bussac, R, Bhudoye, J, Borsarelli
Jazyk: francouzština
Rok vydání: 1989
Předmět:
Zdroj: Annales francaises d'anesthesie et de reanimation. 8(5)
ISSN: 0750-7658
Popis: Three cases of unidirectional valve inversion in airtight oxygen masks are reported. These defects were discovered before use on a patient. One of the masks was tested on a healthy volunteer. It resulted in the rupture of the reservoir bag, with a oxygen leak and decrease in FiO2. Checking the valve before use is therefore mandatory, and fabrication procedures should be revised.