Experience with a higher impedance, fixed helix, steroid-eluting pacing lead. The SWEET PICOTIP Rx Study Group

Autor: M C, Giudici, M, Lee, S, Higgins, G R, West, A K, Moeller, J B, Decker
Rok vydání: 2000
Předmět:
Zdroj: Pacing and clinical electrophysiology : PACE. 23(7)
ISSN: 0147-8389
Popis: A higher impedance, steroid-eluting pacing lead was studied in a multicenter trial. A total of 132 leads were implanted in 61 patients and the electrical characteristics were compared with historical controls. The study lead had a lower pacing threshold at 3 months, 0.72 +/- 0.33 versus 1.29 +/- 0.82 in the atrium and 0.71 +/- 0.25 versus 1.52 +/- 0.52 in the ventricle (P0.01 for both). Impedance was significantly higher in the study lead, 923 +/- 133 versus 719 +/- 109 in the atrium and 980 +/- 235 versus 664 +/- 104 in the ventricle (P0.01 for both). Sensing was similar for both leads. The study lead's higher impedance values and low pacing thresholds combine to reduce pacing system energy requirements.