The effect of computed tomography noise and tissue heterogeneity on cerebral blood flow determination by xenon-enhanced computed tomography

Autor: W F, Good, D, Gur
Rok vydání: 1987
Předmět:
Zdroj: Medical physics. 14(4)
ISSN: 0094-2405
Popis: The errors associated with derivation of cerebral blood flow values by the xenon-enhanced computed tomography (CT) method have been evaluated as a function of tissue heterogeneity and CT noise. The results of this study indicate that CT noise introduces large errors in the derived flow value when data for a single, unprocessed voxel are used for this purpose. CT noise increases the derived flow values in a systematic way. Tissue heterogeneity results in a systematic error which lowers the derived flow values. Errors due to both parameters are computed for typical and extreme conditions.
Databáze: OpenAIRE