Comparing the RETeval

Autor: Jia Yue, You, Allison L, Dorfman, Mathieu, Gauvin, Dylan, Vatcher, Robert C, Polomeno, John M, Little, Pierre, Lachapelle
Rok vydání: 2022
Zdroj: Documenta ophthalmologica. Advances in ophthalmology.
ISSN: 1573-2622
Popis: Our study aimed to determine if ISCEV standard-like ERGs recorded with the LKC RETevalERGs recorded from normal subjects and patients affected with retinal ON and OFF pathway anomalies were compared. Analysis included peak time and amplitude measurements as well as time-frequency domain analysis with the discrete wavelet transform of waveforms obtained with the two systems.Although both systems were similarly able to record reliable and highly reproducible ERG responses, there were major discrepancies in ERG responses between the portable and tabletop units, pointing toward a weaker stimulation of the retinal OFF pathway with the portable RETevalThe portable RETeval