The specific absorption rate (SAR) is a limiting constraint in sequence design for high-field MRI. SAR estimation is typically performed by numerical simulations using generic human body models. This entails an intrinsic uncertainty in present SAR prediction. This study first investigates the required detail of human body models in terms of spatial resolution and the number of soft tissue classes required, based on finite-differences time-domain simulations of a 3 T body coil. The numerical results indicate that a resolution of 5 mm is sufficient for local SAR estimation. Moreover, a differentiation between fatty tissues, water-rich tissues, and the lungs was found to be essential to represent eddy current paths inside the human body. This study then proposes a novel approach for generating individualized body models from whole-body water-fat-separated MR data and applies it to volunteers. The SAR hotspots consistently occurred in the arms due to proximity to the body coil as well as in narrow regions of the muscles. An initial in vivo validation of the simulated fields in comparison with measured B(1)-field maps showed good qualitative and quantitative agreement. |