Popis: |
The purpose of this study was to develop a validated LC-MS/MS analytical method for the simultaneous analysis of a large cassette containing a wide range of drug substances with positive, negative, or neutral charge and further apply the method to assess octanol partition coefficient and eye tissue recovery of the drug cassette. A twenty-seven-drug cassette (N-in-one) including beta blockers, NSAIDs, and corticosteroids that range from extremely hydrophilic (sotalol) to very hydrophobic (triamcinolone hexacetanide) was used to develop an LC-MS/MS assay using QTrap 4500. An LC-MS/MS method based on gradient elution, with an eighteen-minute run time including equilibration time, was developed and validated for the rapid and simultaneous quantitation of drugs with a wide range of lipophilicities. Scheduled multiple reaction monitoring was used to maximize the scan time for each peak, ensuring sufficient scans. Method validation included lower limit of quantitation (LLOQ) and intra- and inter-day reproducibility. The LLOQ ranged from 0.5 (sotalol) to 40 fmols (dexamethasone) on column with a %RSD20%. The method was tested by measuring octanol:water and octanol:buffer (PBS, pH 7.4) partition coefficients and by quantitation of the drug cassette extracted from rabbit aqueous humor and cornea. Measured partition coefficients correlated positively with predicted values (r |